ProductIntroduce£º
Model: 3 1/2 LED
Manual and automatic temperature compensation function
0~10mV DC output signal
Four electrodes: 0.01, 0.1, 1.0 and 10.0cm-1 are available
Automatically switch over measuring frequency. The adoption of titanium alloy
electrode with 0.01cm-1 electrolytic cell constant and seal measuring groove
can meet the measurement of highly pure water.
Specification: 1. Level: Level 1.0
2. Measuring range
Conductivity: (0~1×105)μS/cm
Constant of electrolytic cell matches the most optimum measuring range of conductivity
¡¡¡¡
Constant of electrolytic cell cm-1
¡¡¡¡ 0.01 0.1 1 10
¡¡¡¡
Measuring range μS/cm
¡¡¡¡ 0~2 0.2~20 2~1×104 1×104~1×105
3. Electrode’s constant, cm-1: 0.01, 0.1, 1.0, 10.0
4. Basic error of electronic unit: ±0.5%(FS)±1
5. Stability of electronic unit: ±0.33%(FS)±1/3h
6. Basic error of the instrument: ±1.0%(FS)±1
7. Temperature compensation range: (15~35)°C, base temperature: 25°C
8. External size: 290×200×75mm
9. Weight of the instrument: 1kg
10.Electric power supply: AC220±22V, frequency: 50±1Hz |