Optical Performance
Optical System: Littow type, Single beam, Grating system 1200l/mm
Spectral Slitwidth: 4nm
Wavelength Range:320-1100* nm
Accuracy:±1 nm
Repeatability:0.5 nm
Stray Light:≤0.1%T@340nm
Scan speed: fast, medium, slow
|
Photometric Performance
Photometric Mode: Transmittance,
Absorbance,
Concentration
Photometric Range:0-125.0%T,-0.0970-3.000A, 0-9999C
Accuracy:±0.3%T(0-100%T)
:± 0.002(0-0.5Abs)
:± 0.004(.5-1Abs)
(NIST930)
Repeatability:0.2%T(0-100%T)
0.001(0-0.5Abs)
0.002(0.5-1Abs)
Base line:±0.004A
Drift:±0.002A/h(at 500nm after warm up 1 hour)
Noise :±0.002A
|
Main Functions
Allowed establish standard curve with more than 20
Standards.
Allowed save more than 60 standard curves
100%T:Auto
0%T:Auto
Auto self-test wavelength light source and filter
Data Output: Analog RS-232C
PC Controlled Required: Win-spec.3.33EN Software
Win-spec 3.33EN(option):
Spectrum scanning
Abs/T%/Conc Measurement
Standard curve
Time scanning
|
Specification
Display:2×202-line, 20 character backlighted LCD
Power Requirement: 85-264VAC,50-60Hz
Sample Compartment:100mm light beam for test
Built in 4 position Cell Holder: 10mm
Standard setup of Cells:10mmGlass cell 4 pcs
Dimensions:475 ×420×180(mm)
Net weight :8.5kg
Gross weight:9.5kg
|