ProductIntroduce: Function and Features
l Three kinds of modes: Rate Mode, Time Mode, Body Weight Mode.
l Unique door free structure: Avoiding problems of pump stuck and fluid leakage.
l Double CPU: Making the process of infusion safer and more reliable.
l Accurate Injection: Precise control of the infusion rate and infusion volume.
l Flow rate range: 1.0 ml/h-2000.0 ml/h.
l Driven by step motor: No pulsating wave will be educed even during low-rate
injection.
l Alarm is given in the following situations: Infusion Completion, Occlusion,
Low Battery, Air Bubble, Installation Error, Malfunction or Operation Error.
l IV Set: All brands of standard IV sets are compatible.
l KVO rate can be selectable according to concrete requirements from 1.0 ml/h to 5.0 ml/h.
l Wireless Module Control: When used with SK-2000I or SK-900I, the infusion information will be transferred to the Infusion Supervision System in order to ensure the patients’ safety.
l Big Color LCD.
Technical Parameters of 600II Infusion pump
Dimension
|
120×140×190 mm
|
|
Accumulated infusion volume
|
1.0~9999.0 ml
|
Maximum weight
|
2.0 kg
|
Infusion accuracy
|
±3%
|
Displayed information
|
RATE
VOL.LIMIT
∑( accumulated infusion volume)
Battery charge indicator
AC power indicator
Over
OCCL.
AIR
BED No.
|
Alarm function
|
Over
OCCL.
Low battery
IV line installation error
AIR
EMPTY
Machine malfunction
|
Fuse
|
T 2AL 250V~
|
KVO rate
|
1.0 ml/h-5.0 ml/h
|
Maximum power consumption
|
25VA, running more than 4 hours after being charged
|
Selectable modes
|
Rate mode, Time mode, Weight mode
|
Infusion rate range
|
1.0-2000 ml/h
|
Rate mode
|
1.0~2000.0 ml/h
|
Bolus
|
Rate:≤1000.0 ml/h
Bolus: 1000.0 ml/h
|
Time mode
|
Volume limit: 0.1~999.9ml Time: 1~2000 minutes
|
Operation conditions
|
Environment temperature 5℃~40℃, air pressure 86kPa~106kPa, relative humidity ≤80%
|
Body Weight mode
|
Drug limit: 1.0~999.9 mg, Volume limit: 0.1~999.9 ml. Weight: 0.1~300.0 kg, Dose: 0.1~9999.9
Selectable unit: mg/kg/h, ug/kg/min
|
Electrical Safety
|
Comply with EN60601-1 for medical equipment and EN60601-2-24
|
Battery
|
Rechargeable lithium polymer battery, 7.4V, 1600 mA
|
Applicable infusion sets
|
All standard IV sets
|
Storage
|
Environment temperature-20~55℃, air pressure 50~106kPa, relative humidity ≤95%
|
|