ProductIntroduce: 9-wafer impulse ultrasound transducer
Unique accessory management cabinet
Indication of incorrect manipulation and malfunction analysis
12.1” high resolution color TFT LCD display
Simple and friendly operating display interface
Features & Benefits:
1. 9 standard parameters: ECG, RESP, NIBP, SPO2, 1-TEMP, PR/HR, FHR, TOCO, Fetal movement detection.
2. 12.1” high resolution color TFT LCD display. Simple and friendly operating display interface.
3. 9-wafer impulse ultrasound transducer, configured with special HR self correlation reading algorithm to ensure correct and reliable fetal heart rate (FHR) measurement.
4. Large built-in chargeable battery provides convenience for patients’ moving.
5. Playback and browse function for history waveform and monitor data record.
6. Optional printing output function, alarm triggers printing.
7. Auto double alarm with audible and visible signals.
8. Anti-defibrillation, anti-interference from high frequency electric knife.
9. Pace-maker detection.
10. A full-synchronistic leads, multi-channel ECG display.
11. Indication of incorrect manipulation and malfunction analysis.
12. Unique accessory management cabinet.
Safety
IEC60601-1 approved, Category I BF, CF Type
Dimension and Weight
Dimension: 300mm x 500mm x 350mm
G .Weight: 9.0KGS
Operation Environment
Temperature: Working 5~+40ºC
Humidity: Working ≤80%
Power: AC100-240V, 50/60Hz
Patient range:
Adult, Pediatric, Neonate, and Fetus
Optional:
Thermal Recorder
Performance Specifications
Display: 12.1"color TFT
Rolling and refreshing waveform display
Resolution: 800x600
Trace: 4 waveforms
Sweep speed: 12.5mm/s, 25mm/s, 50mm/s
Indicator: Power/battery indicator light
QRS beep and alarm sound
Battery: Rechargeable built-in battery, above 2 hours for continuous working
Trend time: 72 hours
Alarm: User-adjustable High, Medium and Low limits 3-level Audible and visual alarm
Recorder: Built-in, thermal array
Plethysmogram waveform: 1 channels
Record mode: manual, on alarm, time-defined
Recording width: 50mm
Printing speed: 25mm/s
ECG
Lead mode: 5 –lead, RA, LA, RL, LL, V
lead selection: I, II, III, avR, avL, avF, V, MCL
Sweep speed: 12.5mm/s, 25mm/s, 50mm/s
Heart Rate range: 15~300bpm;
Accuracy: ±1bpm or ±1%, whichever is greater
Resolution: 1bpm
Filter: Surgery mode: 1~25Hz
Monitor mode: 0.5~35Hz
Diagnostic mode: 0.05~100Hz
Scaling signal: 1mV, ±3%
Alarm range: 20~250bpm
S-T segment detection: Measurement range:-2.0mV~ +2.0mV
Alarm Range: -2.0mV~ +2.0mV
NIBP
Operation mode: Manual/Automatic
Auto measurement time: Adjustable (1~120min)
Measurement Unit: mmHg/Kpa selectable
Measurement types: Systolic, Diastolic, Mean
Over-pressure protection: Double safety protection
SPO2
Measurement range: 0~100%
Resolution: 1%
Accuracy: ±2% (80-100%); 1-60% unspecified
Alarm range: 1~100%
RESP:
Method: Thoracic impedance
Measurement range: 1~99rpm
Alarm range: 1~99rpm
Solution: 1rpm
Accuracy: ±2rpm
TEMP
Measurement Range: 25~45 ℃
Resolution: 0.1 ℃
Accuracy: ±0.1 ℃
Channel: 1-channel
PR
Measurement Range: 15~300bpm
Resolution: 1bpm
Accuracy: 1bpm
Alarm range: 15~300bpm
TOCO
Measurement Range 0~100 unit
Non-linearity tolerance ±8%
Accuracy: 1unit
Auto-zero function Yes
FHR
Measurement Range 60~210BPM
Accuracy: ±2rpm
Resolution: 1bpm
Transducer 9-wafer
impulse ultrasound transducer
Ultrasound output intensity ﹤3mW/cm2
Ultrasound working frequency 1.0MHz
Fetal movement record Manual control button label, Auto fetal movement identification
|