ProductIntroduce£º Features
· More accurate
· Safer and more convenient
· CE marked, meet EU clinical requirements
· ISO quality guarantee
· Suitable for adult and pediatric
· Pin-Index for N2O/O2 Cylinder
· Supplementary Supply Port
· 5.7 inch TFT Full-color display
Specifications
Physical and Environmental Specification
1.Demension: L630xW677xH1365
2.Storage temperature: -22°c~+55°c
3.Working temperature: +10°c~+40°c
4.Storage humidity: ≤93°c
Electronic Specification
1.Power supply: AC100~240V±10%, 50/60Hz±1%
2.Power consumption: ≤75VA
3.Top light
Pneumatic Specification
1. Gas supply: O2/N2O/AIR, 0.28MPa~0.6MPa
2. Flowmeters:
O2: 0~1L/min, 1~10L/min
N2O: 0~1L/min, 1~10L/min
AIR: 0~1L/min, 1~10L/min
Six-tube flowmeter, 0~10L/min (With interlock protection devices ensuring the oxygen concerntration > 21%
3. O2 flush: 25~75L/min
4. Working mode: Closed, Semi-closed, Semi-opened
5. APL valve: ≤6KPa
6. O2 supply trouble alarm: O2 pressure lower than 0.14MPa
7. N2O cut off: O2 pressure lower than 0.1MPa
8. Supplementary O2 supply
9. O2/N2O Pin-index
10. Back light for flowmeter
Vaporizer
1. Support 3 vaporizers (Selectatec compatible)
2. Agent type: Sevoflurane, Halothane, Isoflurane available
3. Flow range: 0.2~15L/min
4. Concentration: 0~5% (Sevoflurane 0~8%)
5. Working temerature: 15~35°c
Ventilator specification
1. Patient type: pediatric and adult
2. Ventilation mode: IPPV, IPPB, SIGN, MANUAL
3. Driven mode: Pneumatically driven electronically controled
4. Tidal volume: 50ml~1500ml
5. Ventilation frequency: 0~120bpm
6. I:E rate: 10:1~1:10
7. High pressure: 1~8KPa
8. Low pressure: 0~2KPa
9. Safety valve pressure limit: ≤10KPa
Monitoring
1. Display: high-luminance and wide visual angle TFT LCD display of 5.6 inch
2. Pressure: -2~10KPa
3. Tidal volume: 0~2000ml
4. Respiratory frequency: 0~120bpm
5. Silence time for alarm: ≤120s
6. O2 concentration
7. Waveform: P-T, F-T
Alarm
Upper airway pressure limit
Upper airway pressure limit
Oxygen deficiency |