Power supply
|
Adjustable range
|
With power supply 380V 340V—420V
|
With power supply 220V 195V—250V
|
Fluoroscopy
|
Adjustable range of tube voltage
|
50~95kV (Can adjust continuously)
|
Adjustable range of tube current
|
0.5~5mA (Can adjust continuously)
|
Radiography
|
Adjustable range of tube voltage
|
50~100kV (Can adjust continuously)
|
Adjustable range of tube current
|
Small focus: 50mA
Big focus: 50, 100, 150 and 200mA
|
Tube current of spot-film
|
150mA (fix)
|
Adjustable range of time
|
0.04~6.3s 23 grades together
|
High-voltage
Generator unit
|
Max. DC output voltage
|
100kV
|
Max. DC output current
|
200mA at instant 5mA at running
|
X-ray tube
|
Model
|
XD51-20.40/100
|
Focus
|
1mm×1mm 2mm×2mm
|
Rotary angle
|
+90°~0°~-5°
|
Move range of gastrointestinal spot-film device (fluorescent screen)
|
Pressure direction: ≥300mm
|
Horizontal direction: the center of diagnostic bed moves left (right)
|
≥±80mm
|
Vertical direction: >400mm
|